Conventional rotator cuff versus all-suture anchors-A biomechanical study focusing on the insertion angle in an unlimited cyclic model

  • Dimitris Ntalos
  • Kay Sellenschloh
  • Gerd Huber
  • Daniel Briem
  • Klaus Püschel
  • Michael M Morlock
  • Karl-Heinz Frosch
  • Florian Fensky
  • Till Orla Klatte

Abstract

PURPOSE: The purpose of this study was to compare the biomechanical properties of an all-suture anchor to a conventional anchor used commonly in rotator cuff repairs. Furthermore, the biomechanical influence of various implantation angles was evaluated in both anchor types in a human cadaveric model.

METHODS: 30 humeri were allocated into three groups with a similar bone density. The two different anchor types were inserted at a predefined angle of 45°, 90° or 110°. Biomechanical testing included an initial preload of 20N followed by a cyclic protocol with a stepwise increasing force of 0,05N for each cycle at a rate of 1Hz until system failure. Number of cycles, maximum load to failure, stiffness, displacement and failure mode were determined.

RESULTS: 27 anchors failed by pullout. There was no significant difference between the conventional and the all-suture anchor regarding mean pullout strength. No considerable discrepancy in stiffness or displacement could be perceived. Comparing the three implantation angles no significant difference could be observed for the all-suture or the conventional anchor.

CONCLUSION: All-suture anchors show similar biomechanical properties to conventional screw shaped anchors in an unlimited cyclic model. The exact insertion angle is not a significant predictor of failure.

Bibliografische Daten

OriginalspracheEnglisch
ISSN1932-6203
DOIs
StatusVeröffentlicht - 2019
PubMed 31774856