Ceramic femoral component fracture in total knee arthroplasty: an analysis using fractography, fourier-transform infrared microscopy, contact radiography and histology
Standard
Ceramic femoral component fracture in total knee arthroplasty: an analysis using fractography, fourier-transform infrared microscopy, contact radiography and histology. / Krueger, Alexander P; Singh, Gurpal; Beil, Frank Timo; Feuerstein, Bernd; Ruether, Wolfgang; Lohmann, Christoph H.
in: J ARTHROPLASTY, Jahrgang 29, Nr. 5, 01.05.2014, S. 1001-1004.Publikationen: SCORING: Beitrag in Fachzeitschrift/Zeitung › SCORING: Zeitschriftenaufsatz › Forschung › Begutachtung
Harvard
APA
Vancouver
Bibtex
}
RIS
TY - JOUR
T1 - Ceramic femoral component fracture in total knee arthroplasty: an analysis using fractography, fourier-transform infrared microscopy, contact radiography and histology
AU - Krueger, Alexander P
AU - Singh, Gurpal
AU - Beil, Frank Timo
AU - Feuerstein, Bernd
AU - Ruether, Wolfgang
AU - Lohmann, Christoph H
N1 - Copyright © 2014 Elsevier Inc. All rights reserved.
PY - 2014/5/1
Y1 - 2014/5/1
N2 - Ceramic components in total knee arthroplasty (TKA) are evolving. We analyze the first case of BIOLOX delta ceramic femoral component fracture. A longitudinal midline fracture in the patellar groove was present, with an intact cement mantle and no bony defects. Fractographic analysis with laser scanning microscopy and white light interferometry showed no evidence of arrest lines, hackles, wake hackles, material flaws, fatigue or crack propagation. Analysis of periprosthetic tissues with Fourier-transform infrared (FT-IR) microscopy, contact radiography, histology, and subsequent digestion and high-speed centrifugation did not show ceramic debris. A macrophage-dominated response was present around polyethylene debris. We conclude that ceramic femoral component failure in this case was related to a traumatic event. Further research is needed to determine the suitability of ceramic components in TKA.
AB - Ceramic components in total knee arthroplasty (TKA) are evolving. We analyze the first case of BIOLOX delta ceramic femoral component fracture. A longitudinal midline fracture in the patellar groove was present, with an intact cement mantle and no bony defects. Fractographic analysis with laser scanning microscopy and white light interferometry showed no evidence of arrest lines, hackles, wake hackles, material flaws, fatigue or crack propagation. Analysis of periprosthetic tissues with Fourier-transform infrared (FT-IR) microscopy, contact radiography, histology, and subsequent digestion and high-speed centrifugation did not show ceramic debris. A macrophage-dominated response was present around polyethylene debris. We conclude that ceramic femoral component failure in this case was related to a traumatic event. Further research is needed to determine the suitability of ceramic components in TKA.