AneuGuide™ software-assisted vs. manual measurements in sizing for pipeline embolization device: An agreement study

  • Bin Lv (Geteilte/r Erstautor/in)
  • Mingguang Sun (Geteilte/r Erstautor/in)
  • Yejie Shan (Geteilte/r Erstautor/in)
  • Xiangyu Cao
  • Zhihua Du
  • Xinfeng Liu
  • Rongju Zhang
  • Xiaochang Leng
  • Jens Fiehler
  • Adnan H Siddiqui
  • Jianping Xiang
  • Jun Wang

Abstract

Sizing of flow diverters (FDs) is a challenging task in the treatment of intracranial aneurysms due to their foreshortening behavior. The purpose of this study is to evaluate the difference between the sizing results from the AneuGuide™ software and from conventional 2D measurement. Ninety-eight consecutive patients undergoing pipeline embolization device (PED) treatment between October 2018 and April 2023 in the First Medical Center of Chinese PLA General Hospital (Beijing, China) were retrospectively analyzed. For all cases, the optimal PED dimensions were both manually determined through 2D measurements on pre-treatment 3D-DSA and computed by AneuGuide™ software. The inter-rater reliability between the two sets of sizing results for each methodology was analyzed using intraclass correlation coefficient (ICC). The degree of agreement between manual sizing and software sizing were analyzed with the Bland-Altman plot and Pearson's test. Differences between two methodologies were analyzed with Wilcoxon signed rank test. Statistical significance was defined as p < 0.05. There was better inter-rater reliability between AneuGuide™ measurements both for diameter (ICC 0.92, 95%CI 0.88-0.95) and length (ICC 0.93, 95%CI 0.89-0.96). Bland-Altman plots showed a good agreement for diameter selection between two methodologies. However, the median length proposed by software group was significantly shorter (16 mm versus 20 mm, p < 0.001). No difference was found for median diameter (4.25 mm versus 4.25 mm). We demonstrated that the AneuGuide™ software provides highly reliable results of PED sizing compared with manual measurement, with a shorter stent length. AneuGuide™ may aid neurointerventionalists in selecting optimal dimensions for FD treatment.

Bibliografische Daten

OriginalspracheEnglisch
ISSN0010-4825
DOIs
StatusVeröffentlicht - 01.2024

Anmerkungen des Dekanats

Copyright © 2023 The Authors. Published by Elsevier Ltd.. All rights reserved.

PubMed 38007975