VivaSight Single-Lumen Tube Combined With Hyperangulated Videolaryngoscopy to Rescue Failed Tracheal Intubation in a Patient With Goldenhar Syndrome: A Case Report
Beteiligte Einrichtungen
Abstract
This report describes a patient with Goldenhar syndrome undergoing anesthesia for whom Macintosh videolaryngoscopy failed, as the epiglottis was adhered to the posterior pharynx and could not be lifted with a tracheal introducer (Cormack-Lehane grade 3B). Hyperangulated videolaryngoscopy revealed only the arytenoids (Cormack-Lehane grade 2B), even after direct lifting of the epiglottis, and endotracheal tube advancement failed due to unclear tissue resistance. Hyperangulated videolaryngoscopy was combined with a tube-mounted camera (VivaSight single lumen tube). The combination of both camera perspectives was successfully used to allow placement of the endotracheal tube underneath the epiglottis and through the vocal cords.
Bibliografische Daten
Originalsprache | Englisch |
---|---|
Aufsatznummer | e01615 |
ISSN | 2575-3126 |
DOIs | |
Status | Veröffentlicht - 01.09.2022 |
Anmerkungen des Dekanats
Copyright © 2022 International Anesthesia Research Society.
PubMed | 36099152 |
---|