Miniplate osteosynthesis with four different systems in sheep

  • R Gutwald
  • R Schön
  • M Metzger
  • K Kreutzer
  • B Rahn
  • R Schmelzeisen
  • S Sauerbier

Abstract

The aim of this study was to compare a combination of a locking system with self-tapping (ST-L) or self-drilling-tapping (SDT-L) screws with a combination of conventional miniplates with self-tapping (ST) and self-forming (SF) screws. A standardized osteotomy and osteosynthesis with one of the above mentioned systems was performed in 24 sheep. Callus formation was measured with the help of CT scans assisted by a navigation system. Specimens of each osteotomy gap were taken and examined histologically. The best results were observed when self-tapping screws and the Mini-Locking-System (ST-L) were applied. The slowest healing was seen in animals treated with miniplates and SF screws. After 8 weeks an increase in bone formation could be observed in the ST, SF, SDT-L systems. The results after 8 weeks were comparable with those achieved by the ST-L system after 4 weeks. The improved stability of the osteosynthesis with the ST-L system resulted in early ossification of the osteotomy gap and the smallest amount of callus formation.

Bibliografische Daten

OriginalspracheEnglisch
ISSN0901-5027
DOIs
StatusVeröffentlicht - 01.2011
Extern publiziertJa
PubMed 21084173